Real-world application of an integrated MCDA–economic framework in continence care technology: a hospital-based case study

Authors

  • Gianpaolo Franzoso Pharmacy unit, Veneto Institute of Oncology IOV-IRCCS, Padua - Italy https://orcid.org/0009-0004-3804-3556
  • Novella Piazzetta Pharmacy unit, Veneto Institute of Oncology IOV-IRCCS, Padua - Italy https://orcid.org/0009-0001-5425-8539
  • Giorgia Zorzetto Pharmacy unit, Veneto Institute of Oncology IOV-IRCCS, Padua - Italy https://orcid.org/0000-0002-2528-691X
  • Mirsad Pasalic Medical Oncology inpatient unit, Veneto Institute of Oncology IOV-IRCCS, Padua - Italy
  • Valentina Ceron Medical Oncology inpatient unit, Veneto Institute of Oncology IOV-IRCCS, Padua - Italy
  • Flavia Damiano Surgical Inpatient Unit Nurse Manager, Veneto Institute of Oncology IOV-IRCCS, Padua - Italy https://orcid.org/0009-0002-3229-953X
  • Cristina Tridello Surgical Inpatient Unit Nurse Manager, Veneto Institute of Oncology IOV-IRCCS, Padua - Italy
  • Marina Coppola Pharmacy unit, Veneto Institute of Oncology IOV-IRCCS, Padua - Italy

DOI:

https://doi.org/10.33393/grhta.2026.3828

Keywords:

Budget impact analysis, Cost-consequence analysis, Continence care, Hospital-based HTA, Multi-criteria decision analysis, Medical devices, Real-world evidence analysis

Abstract

Introduction: Hospital-based Health Technology Assessment (HTA) supports transparent decisions about adopting medical technologies by balancing practical benefits against clinical, organizational, and economic consequences. This study evaluated a single-use continence care system using an integrated decision-support framework.
Methods: The single-use system, based on disposable absorbent components, was compared with the standard reusable bedpan/urinal pathway, including washing and reprocessing. The assessment combined Multi-Criteria Decision Analysis (MCDA), Budget Impact Analysis (BIA), Cost-Consequence Analysis (CCA), a 30-day field test across inpatient units at two hospital sites, and an implementation-specific questionnaire. During the field test, 168 admissions were recorded. MCDA used a weighted aggregate score, while the economic analysis adopted the hospital perspective over three years. Deterministic sensitivity analyses examined economic inputs and MCDA domain weights.
Results: The single-use system offered perceived advantages in ease of use, hygiene, and odor reduction. However, these did not offset limited supporting evidence, lower patient-comfort ratings, operational issues, and higher costs. The MCDA score decreased from 4.80 for standard care to 2.60 for the single-use system (difference: −2.20). Cost per patient increased from €0.65 to €1.00, an incremental cost of €0.35. The three-year budget impact was €2,634 and remained within the local affordability threshold. As the technology was more costly and produced lower overall MCDA value, it was classified as dominated. Sensitivity analyses did not alter the recommendation.
Conclusion: The integrated MCDA-economic framework supported transparent hospital-level decision-making. Affordability and perceived practical advantages were insufficient to justify routine adoption of the single-use continence care system.

Introduction

Hospital-based Health technology assessment (HTA) plays a key role in supporting decisions on the adoption, replacement or disinvestment of medical technologies. Unlike national or regional HTA processes, hospital-based HTA is closely linked to local workflows, available resources, procurement constraints and organizational priorities. For this reason, it requires methods that are sufficiently structured to ensure transparency, while remaining adaptable to the operational context in which the technology will be used (1-4).

The implementation of Regulation (EU) 2021/2282 has strengthened the role of HTA within European health technology governance, particularly by promoting common clinical assessment and methodological convergence across Member States (2,3). At the same time, hospital-level decisions continue to require local information on workflow, resource use, implementation feasibility and budget impact that may not be fully captured by supra-national or national assessments (4-6).

In this context, decision-making may be particularly challenging when a technology offers intuitive or perceived advantages, such as greater ease of use, improved hygiene or reduced handling burden, while robust comparative clinical evidence is limited. If such technologies are evaluated only through a narrow economic lens, relevant organizational or experiential consequences may be missed. Conversely, if they are evaluated only through user perception or qualitative preference, their real cost and operational impact may be underestimated.

Multi-Criteria Decision Analysis (MCDA) has been increas-ingly used in healthcare to structure complex decisions involving multiple dimensions of value. MCDA allows decision-makers to combine heterogeneous criteria, such as clinical benefit, safety, organizational consequences, patient-related outcomes, and technical relevance, through an explicit scoring and weighting process (7,8).

To address this need, an institutional decision-support framework was developed to integrate MCDA with structured economic evaluation. The framework combines a value-risk appraisal with Budget Impact Analysis (BIA), Cost-Consequence Analysis (CCA), and a rule-based recommendation matrix, allowing decision-makers to assess both multidimensional value and economic sustainability within a single process (7-10).

Previously published hospital-based HTA decision-support tools and prioritization frameworks have emphasized the need for multidimensional criteria, stakeholder-informed scoring, and explicit consideration of local implementation conditions (6,11). However, economic assessment is often reported as a parallel component or summarized as a cost criterion within MCDA, rather than being operationally connected to field-testing evidence, BIA, CCA, and a final recommendation rule. The contribution of this case study is therefore not the proposal of a new generic MCDA theory, but the transparent application of an integrated local workflow that links real-world ward feedback, weighted MCDA, budget impact, cost-consequence interpretation, and dominance-based decision rules within a single hospital adoption decision.

This study presents a real-world application of the framework to the evaluation of a continence care technology in a hospital inpatient setting. The assessed technology was a single-use system intended to replace, at least partially, standard reusable bedpans and urinals. The case was selected because it illustrates a frequent hospital decision problem: a technology with plausible practical benefits and positive user perception, but uncertain incremental value when clinical, economic, and organizational dimensions are assessed together.

Methods

Study design and setting

A hospital-based HTA case study was conducted using an integrated MCDA-economic framework. The evaluation was performed in an oncological hospital with two sites, but the decision problem was not specific to oncology: the target population was represented by inpatients with reduced mobility requiring assisted toileting or excreta management. The assessment was designed to support an institutional decision regarding the potential adoption of the technology in routine care.

The structure of the assessment was consistent with the multidimensional perspective recommended in HTA models and with good-practice guidance for healthcare MCDA and BIA (4,7-10). The framework combined quantitative scoring, local economic estimates, field testing and qualitative observations in order to capture both measurable outcomes and implementation-related consequences.

Intervention, comparator and PICO

The intervention assessed was a single-use continence care system based on disposable absorbent components for excreta collection. The technology is intended to simplify the management of patient elimination needs by avoiding part of the handling, washing, and reprocessing activities associated with reusable devices.

The comparator was the standard reusable bedpan/urinal pathway currently used in routine care. This pathway includes the use of reusable devices and the related activities required for collection, transport, washing, sanitization, and reprocessing. The comparison therefore considered not only acquisition costs, but also the broader care pathway and the organizational consequences of each option.

The PICO was defined as follows: Population, patients with reduced mobility requiring assisted toileting; Intervention, disposable excreta collection kit; Comparator, reusable bedpan/urinal pathway; Outcome, reduction in potential cross-contamination risk, assessed through documented reprocessing requirements, staff-reported operational issues, and availability of evidence supporting infection-control benefit.

Field testing and questionnaire-based data collection

Field testing was conducted for 30 consecutive days in inpatient areas across the two hospital sites. The technology was used in routine care for non-self-sufficient inpatients requiring assisted toileting or excreta management. During the 30-day observation period, 168 admissions were recorded in the involved inpatient areas. This figure was used as the pragmatic field-test denominator for the clinical context; no formal sample-size calculation was performed because the study was designed as a hospital-based implementation assessment rather than as an effectiveness trial.

Structured questionnaires were administered to healthcare professionals involved in the use of the technology, including nursing staff, nursing coordinators and healthcare assistants. The questionnaire was developed specifically for this implementation assessment by the institutional HTA team and was not adapted from a previously validated instrument. It used a 1-5 increasing positivity scale and covered current practice, perceived safety, practical benefits, operator workload, patient-care experience, and perceived quality of care. The items were designed to mirror the organizational, technical, patient-related, and safety dimensions used in the MCDA and CCA. No formal psychometric validation or inferential survey analysis was performed; the questionnaire was used as a pragmatic tool to support real-world implementation appraisal. Free-text and informal comments collected during ward visits were used to contextualize the structured responses.

The pharmacist responsible for the assessment regularly visited the participating wards to collect completed questionnaires, verify practical issues with operators, and discuss implementation problems directly with the staff. Questionnaire responses and qualitative observations were not used as stand-alone evidence of clinical effectiveness; they supported the scoring of organizational, technical, and patient-related criteria and informed the CCA interpretation.

MCDA methodology

The MCDA component assessed the value-risk profile of the technology across domains relevant to hospital-based HTA. The criteria reflected the institutional framework aligned with the regional MCDA logic and included clinical effectiveness, safety, organizational impact, patient experience, and economic impact. Criteria and weights were not elicited ad hoc for this single case study. They were derived from a predefined institutional framework aligned with the regional RATEC approach, in which criteria and relative weights had been defined through a structured expert-consensus process involving members of Medical Device Assessment Units and multidisciplinary hospital stakeholders, including clinical, pharmaceutical, nursing, technical, and managerial/procurement perspectives. The weighting process was consensus-based rather than a formal AHP or Delphi exercise, and the resulting weights were adopted as fixed inputs for this case-study application. For transparency and reproducibility, all aggregate weights used in the present assessment are reported in Table 1, while the subcriterion structure used for the value-risk dashboard is reported in Table 2.

For each criterion, the new technology was compared with the current standard and assigned a score on a 0-1-3-5-7-9 ordinal scale, where higher values indicate a more favorable performance. Evidence certainty and contextual validity influenced scoring through structured critical appraisal. When evidence was indirect or limited, scores were downgraded even when a plausible benefit was identified. Field-testing results were used only for criteria in which real-world implementation experience was directly relevant, such as workflow, usability, training needs, and patient-care experience (Tables 1 and 2).

The weighted aggregate MCDA score was calculated as MCDA = Σ(wj × Sj), where wj is the predefined weight of criterion j, and Sj is the score assigned to the technology for that criterion. The incremental MCDA value was calculated as ΔMCDA = MCDAnew - MCDAstandard. Positive ΔMCDA values indicate a higher value of the new technology compared with the comparator, whereas negative values indicate a lower value.

The value-risk plot was used as a complementary graphical output. It summarizes the position of the new technology in a matrix based on value and risk coordinates. These coordinates are related to, but not numerically identical to, the weighted aggregate MCDA score because they are generated from the value-risk decomposition used by the institutional dashboard. Final interpretation was therefore based on the integrated rule-based decision matrix rather than on the graph alone.

Economic evaluation

The economic evaluation was conducted from the hospital provider perspective. The time horizon for the BIA was 3 years, consistent with the institutional decision process. Costs were expressed in euros and were based on internal acquisition prices, local consumption estimates, and assumptions on resource use for the reusable pathway and the single-use system. No discounting was applied because the analysis was a short-term budget impact assessment rather than a long-term cost-effectiveness model.

The costing approach distinguished common pathway components from incremental decision-relevant costs. Some resources, such as staff handling time or basic protective materials, were present in both pathways and therefore had limited impact on the incremental comparison. The economic model used the relevant incremental cost per patient for the BIA: €0.65 for the reusable standard pathway and €1.00 for the single-use system, corresponding to ΔCost = €0.35 per patient. The model also included a one-off implementation/acquisition cost of €1,500 in the first year.

Criterion Weight Standard score New score Standard weighted score New weighted score Δ Main rationale
Clinical effectiveness 0.10 3 5 0.30 0.50 +0.20 Marginal potential benefit; no robust comparative clinical evidence
Safety 0.15 3 5 0.45 0.75 +0.30 Theoretical reduction in cross-contamination risk; evidence indirect
Organizational impact 0.30 5 3 1.50 0.90 -0.60 Workflow issues, training needs, and bag-management problems observed in field testing
Patient experience 0.25 3 1 0.75 0.25 -0.50 Reduced comfort and bedside management concerns reported by operators
Economic impact 0.20 9 1 1.80 0.20 -1.60 Higher cost per patient for the single-use system
Total weighted aggregate score 1.00 4.80 2.60 -2.20 Overall lower value of the new technology
Table 1 -. Complete aggregate MCDA scoring matrix used in the case study

Budget impact was calculated by comparing a baseline scenario in which all eligible patients remained on the standard pathway with a mixed scenario in which the single-use system was introduced according to predefined adoption rates. CCA was used to present costs and consequences separately across clinical, organizational, patient-related, and economic dimensions. The simplified cost-value indicator previously labeled as “soft ICER” was retained only as a supportive internal cost-per-MCDA-unit indicator. It is not comparable with conventional ICERs based on QALYs and is not interpreted when a technology is dominated (Table 3).

Domain Domain weight Subcriterion Impact Evidence score Subweight Index Rationale
Technical relevance 0.20 Operator skills/professional expertise −1 7 0.24 −0.0806 Additional operator requirements
Technical relevance 0.20 Ease/comfort of use −1 5 0.18 −0.0324 Practical issues despite perceived ease in selected steps
Patient benefit 0.50 Quality of life −1 3 0.26 −0.1014 Reduced patient comfort/bedside management concerns
Patient benefit 0.50 Specific clinical outcomes +1 3 0.15 +0.0338 Potential but limited outcome relevance
Patient benefit 0.50 Risk of complications +1 3 0.22 +0.0726 Theoretical reduction in cross-contamination risk
Economic impact 0.42 Economic impact domain +1 risk contribution 9 1.00 +3.7800 Higher cost increases the risk component
Organizational impact 0.58 Medical team working time +1 risk contribution 5 0.19 −0.1047 Expected time savings not fully realized
Organizational impact 0.58 Healthcare performance +1 risk contribution 5 0.38 −0.4188 Workflow and implementation concerns
Table 2 -. Value-risk subcriterion matrix used to position the technology in the MCDA dashboard. Positive and negative signs reflect the value-risk scoring logic of the dashboard and should be interpreted together with the final recommendation

figure 1. - Overall decision framework showing the interaction between field testing, MCDA, BIA, CCA and the rule-based recommendation.

Uncertainty analysis

To assess uncertainty, deterministic one-way sensitivity analyses were performed on the main parameters most likely to influence the recommendation: the cost of the new technology, the cost of the standard pathway, adoption volumes, and the MCDA score of the new technology. In addition, a simple MCDA structural sensitivity analysis was added to explore uncertainty in domain weights. The tested scenarios increased the weight of the potentially favorable domains (clinical effectiveness and safety) by 20%, decreased the economic-impact weight by 20%, and decreased the organizational and patient-experience weights by 20%, with proportional rescaling of the remaining domains to keep the total weight equal to 1.00. These scenarios were selected to test whether plausible changes in the most influential domains could reverse the dominance interpretation. A probabilistic sensitivity analysis was not performed because the assessment relied on a single-center implementation dataset and local estimates rather than on parameter distributions derived from comparative trials.The sensitivity analyses were interpreted according to whether they changed the dominance status or the final rule-based recommendation (Table 4).

Decision rule

The final recommendation was generated through a rule-based integration of the MCDA output and economic results. First, the MCDA value-risk profile identified whether the new technology provided additional value and whether relevant risks were present. Second, the BIA assessed affordability within local budget thresholds. Third, the CCA described the direction of consequences across clinical, organizational, patient-related, and economic dimensions. Finally, the supportive cost-per-MCDA-unit indicator was considered only when a meaningful trade-off existed.

When the new technology generated a lower MCDA value and a higher cost than the comparator, it was classified as dominated. In this situation, the cost-per-MCDA-unit indicator was considered not applicable, because calculating a ratio would not add meaningful information to the decision.

Results

Field testing and questionnaire findings

During the 30-day observation period, 168 admissions were recorded in the involved inpatient areas. Healthcare professionals reported perceived advantages related to ease of use in selected steps, perceived hygiene, and odor reduction. These findings suggested that the technology addressed practical needs that can be meaningful in routine inpatient care.

At the same time, several critical issues emerged during field testing. The single-use system did not produce a consistent reduction in procedure time, required specific training, and introduced practical complexity in bag management, including closure, emptying, and disposal. In some settings, keeping the device at the bedside generated concerns about space management, perceived hygiene, and patient comfort. These observations were central to the final appraisal because they showed that the theoretical simplification of the care pathway did not fully materialize in real-world use (Table 5).

Parameter Value/assumption
Analytical perspective Hospital provider perspective
Currency Euro (€)
Time horizon 3 years
Data sources Internal acquisition prices, local consumption estimates, field-test observations, ward feedback and decision-support application outputs
Cost per patient - reusable standard pathway €0.65
Cost per patient - single-use system €1.00
Incremental cost per patient €0.35
One-off implementation/acquisition cost €1,500 in year 1
Adoption scenario 50% in year 1, 60% in year 2, 90% in year 3
Target population volumes 1,500 patients in year 1; 1,600 in year 2; 1,700 in year 3
Discounting Not applied, because the analysis was a short-term BIA
Cost-value indicator Supportive cost-per-MCDA-unit indicator; not applicable when the new technology is dominated
Table 3 -. Economic evaluation assumptions and input parameters
Scenario Parameter value Economic result ΔMCDA Interpretation
New technology cost per patient -10% €0.90 +€0.25 -2.20 Dominated; not favorable
New technology cost per patient +10% €1.10 +€0.45 -2.20 Dominated; not favorable
Standard pathway cost per patient -10% €0.585 +€0.415 -2.20 Dominated; not favorable
Standard pathway cost per patient +10% €0.715 +€0.285 -2.20 Dominated; not favorable
Patients treated with new technology -10% 2,916 over 3 years 3-year BIA €2,520.60 -2.20 Within threshold; not favorable
Patients treated with new technology +10% 3,564 over 3 years 3-year BIA €2,747.40 -2.20 Within threshold; not favorable
New technology MCDA score +10% 2.86 +€0.35 -1.94 Dominated; not favorable
Clinical effectiveness and safety weights +20%; other domains proportionally rescaled Clinical effectiveness 0.12; safety 0.18 +€0.35 -1.92 Dominated; not favorable
Economic-impact weight -20%; other domains proportionally rescaled Economic impact 0.16 +€0.35 -1.91 Dominated; not favorable
Organizational and patient-experience weights -20%; other domains proportionally rescaled Organizational 0.24; patient experience 0.20 +€0.35 -2.25 Dominated; not favorable
Table 4 -. Deterministic sensitivity analysis, including economic and MCDA domain-weight scenarios
Positive findings Critical findings
Ease of use in selected steps No consistent reduction in overall procedure time
Improved perceived hygiene Complexity in bag closure, emptying, and disposal
Reduction of odors Need for specific training and initial learning phase
Positive staff perception in selected situations Concerns about bedside device management and patient comfort
Potential use in specific high-infectious-risk contexts Routine adoption not supported without procedure revision and further evidence
Table 5 -. Summary of field-testing findings

MCDA results

The MCDA assessment showed a lower overall value-risk score for the single-use system compared with standard care. The current reusable pathway obtained a weighted aggregate MCDA score of 4.80, whereas the single-use system obtained a score of 2.60, corresponding to ΔMCDA = -2.20.

The negative result was mainly driven by the economic impact of the single-use system, patient-comfort concerns, and organizational issues observed during field testing. Potential advantages were identified in safety, mainly related to theoretical reduction of cross-contamination risk, but the supporting evidence was indirect and did not compensate for the negative findings in other domains. The value-risk plot positioned the technology in the unfavorable area, with a risk score of 5.43 and value score of 0.08 (Fig. 2).

Economic and budget impact results

The single-use system was associated with a higher cost per patient than the standard reusable pathway. The cost per patient was €1.00 for the single-use system and €0.65 for the reusable standard, resulting in an incremental cost of €0.35 per patient (Table 6).

The BIA estimated a year-1 budget impact of €1,762.50 and a 3-year cumulative budget impact of €2,634. Although this amount remained within the institutional affordability threshold, the incremental expenditure was not justified by a corresponding increase in multidimensional value. The BIA therefore did not reverse the unfavorable interpretation generated by the MCDA and CCA components.

CCA and final integrated decision

The CCA showed a mixed profile. The single-use system improved perceived safety in relation to a theoretical reduction of cross-contamination risk. However, it worsened patient comfort, operator time, training requirements, organizational impact, and direct costs. Several dimensions were neutral, including clinical effectiveness, procedure time, space or instrumentation, indirect costs, and recommended use.

Because the technology was associated with higher incremental cost and lower incremental MCDA value, it was classified as dominated. In this situation, the supportive cost-per-MCDA-unit indicator was not informative and was considered not applicable. The final rule-based recommendation was therefore not favorable for routine adoption. Selective use could be considered only in clearly defined subgroups or settings, such as patients at particularly high infectious risk, and should be accompanied by additional evidence collection, staff training, and reassessment of operational procedures.

figure 2 -. Value-risk MCDA results for the Kit Hygie case study. The new technology is located in the unfavorable zone, characterized by a high risk (5.43) paired with a low value (-0.08).

Description Unit Cost (€) Total cost new technology (€) Total cost comparator (€)
Gloves 0.03 0.03 0.03
Underpad 40 × 60 cm 0.15 - 0.15
Bedpan 4.67 - 0.05
Washing cycle - - 0.45
Healthcare Assistant (HCA) hourly rate 0.42 1.26 1.26
Hygie Kit 1.00 1.00 -
TOTAL - 2.29 1.94
Table 6 -. Cost comparison per procedure between the disposable system and the traditional bedpan pathway

figure 3 -. Although the three-year BIA shows sustainability, the introduction of the new technology is unfavorable as it is associated with higher costs and lower value.

Sensitivity analysis results

The deterministic sensitivity analyses did not change the final recommendation. Across all tested economic scenarios, the single-use system remained more costly than the comparator or retained a lower MCDA value. Even when the new technology score was increased by 10%, ΔMCDA remained negative (-1.94). The MCDA domain-weight scenario analysis also did not reverse the conclusion: increasing the weights of clinical effectiveness and safety by 20% resulted in ΔMCDA = -1.92, and reducing the economic-impact weight by 20% resulted in ΔMCDA = -1.91. The conclusion was therefore robust to plausible one-way variations in both economic inputs and principal MCDA domain weights.

Discussion

This case study illustrates the practical value of an integrated MCDA-economic framework for hospital-based HTA. The evaluated technology had intuitive advantages and generated some positive perceptions among healthcare professionals. However, the structured assessment showed that these perceived benefits were not sufficient to produce a favorable overall value-risk profile when evidence, organizational impact, and economic consequences were considered together (7-10).

The most relevant finding is the discrepancy between perceived usefulness and overall measurable value. In many hospital decisions, technologies may appear attractive because they simplify a visible part of the care process or address an operational discomfort. Nevertheless, the overall value of a technology depends on the balance between benefits, evidence, costs, workflow impact, and feasibility. In the present case, the single-use system increased cost per patient and did not improve the overall MCDA value-risk profile. The integrated framework therefore prevented a decision based solely on subjective perception or isolated operational advantages.

figure 4 -. Results of the CCA for the Kit Hygiene. The dashboard highlights a predominantly negative (worsens) or neutral (neutral) impact across key organizational, direct cost, and perceived quality dimensions, contrasted with a single theoretical improvement (improves) in safety.

The classification of the technology as dominated is particularly informative. A purely economic evaluation would have shown that the single-use system was more costly than the reusable pathway, but it would not have explained whether the additional cost could be justified by broader dimensions of value. Conversely, a purely qualitative assessment might have emphasized perceived hygiene and usability without adequately accounting for cost and organizational burden. The integrated framework made these trade-offs explicit.

Another important aspect is the role of field testing. Product characteristics and theoretical pathway advantages are often insufficient to predict real-world performance. In this case, field observations showed that the expected simplification of the process was limited by bag management issues, training needs, and workflow adaptation. These findings support the inclusion of implementation data in hospital-based HTA, especially for technologies whose value depends on routine use by healthcare professionals.

The supportive cost-per-MCDA-unit indicator should be interpreted with caution. The term ‘soft ICER’ was used internally to describe a pragmatic ratio between incremental cost and incremental MCDA value, but it should not be confused with conventional ICERs based on QALYs or other health outcomes. Because MCDA scores are dimensionless and institution-specific, this indicator cannot be compared across technologies or settings without recalibration. Moreover, when a technology is dominated, as in this case, calculating a ratio is unnecessary and potentially misleading.

The study also shows the usefulness of CCA and BIA in combination with MCDA. CCA allowed costs and consequences to remain visible as distinct components rather than being compressed into a single economic endpoint. BIA supported the interpretation of affordability and local expenditure consequences, which are essential for hospital-level decisions where budget constraints and implementation feasibility strongly influence adoption (9,10).

This case has practical implications for hospital decision-makers. First, technologies with weak evidence and mainly perceived benefits should be assessed with particular attention to organizational consequences. Second, small differences in unit cost may become relevant when the incremental value is limited or negative. Third, structured frameworks can improve accountability by making explicit why a technology is adopted, rejected, or reserved for selected contexts.

Generalizability and transferability

Although the assessment was conducted in a single institution, the clinical use case is not specific to oncology. Assisted toileting and excreta management for non-self-sufficient inpatients occur across many hospital settings, including medical, surgical, rehabilitation, and long-term care wards. Therefore, the decision problem is potentially transferable beyond the oncological setting.

The most transferable components of the framework are the decision structure, the explicit MCDA scoring logic, the BIA/CCA integration, the dominance interpretation, and the rule-based recommendation hierarchy. Conversely, several inputs are context-specific and should be recalibrated before application elsewhere: local reprocessing costs, staff time, washer-disinfector availability, procurement prices, adoption volumes, waste-management procedures, infection-control protocols, and institutional affordability thresholds.

In hospitals where reprocessing is outsourced or associated with higher operational costs, the economic profile of single-use solutions may become more favorable. Conversely, in hospitals with efficient in-house reprocessing and established workflows, the incremental value of single-use systems may be lower. The framework is therefore generalizable as a method, whereas the case-study result should be interpreted as context-specific.

Limitations

The assessment has limitations. First, it was conducted in a single hospital and was based on local costs, workflows, and professional feedback. Second, field testing lasted 30 days and may not fully reflect longer-term adaptation by staff. Third, the questionnaire was developed specifically for this implementation assessment and was not formally validated or pilot-tested as a psychometric instrument; it should therefore be interpreted as a pragmatic data-collection tool rather than as a validated measurement scale. Fourth, the number of questionnaire responses was not used to estimate a formal response rate, because the assessment was designed as a pragmatic implementation study rather than as a survey with inferential objectives. Fifth, the weighting system was derived from a predefined consensus-based institutional/regional framework and was not re-elicited for this individual case. Although domain-weight sensitivity analysis did not change the recommendation, future applications should further evaluate weight stability across stakeholders and settings. Finally, evidence supporting infection-control benefit was indirect and did not demonstrate a robust clinical advantage, and probabilistic sensitivity analysis was not feasible because parameter distributions were not available.

Despite these limitations, the case study provides a useful example of how hospital-based HTA can integrate multidimensional value and economic sustainability in a reproducible decision process. The findings are consistent with the broader methodological direction of HTA and MCDA guidance, which emphasizes transparent criteria, explicit trade-offs, and the integration of contextual factors into decision-making (1,4,7,8).

Conclusions

The integrated MCDA-economic framework provided a transparent and structured approach to the hospital-level evaluation of a single-use continence care technology. In the present case, the technology showed perceived advantages in hygiene, odor reduction and ease of use, but these benefits were not supported by sufficient evidence or accompanied by improved overall value.

Compared with the reusable pathway, the single-use system was associated with higher cost per patient and a lower MCDA score. The technology was therefore classified as dominated within the integrated appraisal and did not demonstrate sufficient incremental value to justify routine adoption.

This case supports the use of integrated hospital-based HTA frameworks to guide decisions on medical technologies, especially when perceived practical benefits must be balanced against evidence uncertainty, organizational consequences and economic sustainability.

Other information

Corresponding author:

Giorgia Zorzetto

email: giorgia.zorzetto@iov.veneto.it

Disclosures

Conflict of interest: The authors declare no conflict of interest.

Financial support: This research received no specific grant from any funding agency in the public, commercial, or not-for-profit sectors.

Data availability statement: The data used for the assessment were derived from local institutional sources and the decision-support application. Aggregated data relevant to the interpretation of the case study are reported in the manuscript.

References

  1. Ferrante di Ruffano L, Harris IM, Zhelev Z, et al. Health technology assessment of diagnostic tests: a state of the art review of methods guidance from international organizations. Int J Technol Assess Health Care. 2023;39(1):e14. https://doi.org/10.1017/S0266462323000065 PMID:36803886
  2. European Parliament and Council of the European Union. Regulation (EU) 2021/2282 of December 15, 2021 on health technology assessment and amending Directive 2011/24/EU. Official Journal of the European Union. 2021;L458:1-32. Available from: https://eur-lex.europa.eu/eli/reg/2021/2282/oj (Accessed May 2026)
  3. European Commission. Implementation of the Regulation on health technology assessment. Brussels: European Commission; Available from: https://health.ec.europa.eu/health-technology-assessment/implementation-regulation-health-technology-assessment_en [accessed 28 September 2026]. (Accessed May 2026)
  4. EUnetHTA Joint Action 2. HTA Core Model® version 3.0. 2016. Available from: https://www.htacoremodel.info/model/HTACoreModel3.0.pdf (Accessed May 2026)
  5. Agenzia Nazionale per i Servizi Sanitari Regionali (AGENAS). Programma Nazionale HTA Dispositivi Medici 2023-2025. AGENAS; 2024.
  6. Sampietro-Colom L, Morilla-Bachs I, Gutierrez-Moreno S, et al. Development and test of a decision support tool for hospital health technology assessment. Int J Technol Assess Health Care. 2012;28(4):460-465. https://doi.org/10.1017/S0266462312000487 PMID:23062518
  7. Thokala P, Devlin N, Marsh K, et al. Multiple criteria decision analysis for health care decision making—an introduction: report 1 of the ISPOR MCDA emerging good practices task force. Value Health. 2016;19(1):1-13. https://doi.org/10.1016/j.jval.2015.12.003 PMID:26797229
  8. Marsh K, IJzerman M, Thokala P, et al.; ISPOR Task Force. Multiple criteria decision analysis for health care decision making—emerging good practices: report 2 of the ISPOR MCDA emerging good practices task force. Value Health. 2016;19(2):125-137. https://doi.org/10.1016/j.jval.2015.12.016 PMID:27021745
  9. Sullivan SD, Mauskopf JA, Augustovski F, et al. Budget impact analysis-principles of good practice: report of the ISPOR 2012 Budget Impact Analysis Good Practice II Task Force. Value Health. 2014;17(1):5-14. https://doi.org/10.1016/j.jval.2013.08.2291 PMID:24438712
  10. Drummond MF, Sculpher MJ, Claxton K, et al. Methods for the Economic Evaluation of Health Care Programmes. 4th ed. Oxford University Press; 2015.
  11. Martin J, Lal A, Moodie J, et al. Hospital-Based HTA and Know4Go at MEDICI in London, Ontario, Canada. In: Sampietro-Colom L, Martin J, eds. Hospital-Based Health Technology Assessment: The Next Frontier for Health Technology Assessment. Springer; 2016:127-152, https://doi.org/10.1007/978-3-319-39205-9_12.
  12. Ruggeri M, Cadeddu C, Roazzi P, et al. Multi-Criteria-Decision-Analysis (MCDA) for the horizon scanning of health innovations an application to COVID 19 emergency. Int J Environ Res Public Health. 2020;17(21):7823. https://doi.org/10.3390/ijerph17217823 PMID:33114587

Author Biographies

Gianpaolo Franzoso, Pharmacy unit, Veneto Institute of Oncology IOV-IRCCS, Padua - Italy

Gianpaolo Franzoso is Head of the Simple Operating Unit for Medical Device and Health Technology Governance at the Veneto Institute of Oncology IOV-IRCCS, Padua, Italy. He also serves as RLV and President of the UVA-DM at the same institution.

Novella Piazzetta, Pharmacy unit, Veneto Institute of Oncology IOV-IRCCS, Padua - Italy

Novella Piazzetta is a scholarship holder at the Pharmacy Unit of the Veneto Institute of Oncology IOV-IRCCS, Padua, Italy.

Mirsad Pasalic, Medical Oncology inpatient unit, Veneto Institute of Oncology IOV-IRCCS, Padua - Italy

Dr. Mirsad Pasalic is the Nursing Coordinator for the Medical Inpatient Area at the Veneto Institute of Oncology IOV – IRCCS, Padua, Italy. In this role, he is involved in the organization and coordination of nursing activities within medical inpatient wards, contributing to the management of clinical care pathways and multidisciplinary collaboration in an oncology hospital setting.

Valentina Ceron, Medical Oncology inpatient unit, Veneto Institute of Oncology IOV-IRCCS, Padua - Italy

Dr. Valentina Ceron holds a Functional Role Appointment in the Medical Area, with responsibility for Oncology 3, Oncohematology and Hospice at the Veneto Institute of Oncology IOV – IRCCS, Castelfranco Veneto, Italy. In this role, she contributes to the organization and coordination of care activities within oncology, oncohematology and palliative care settings, supporting multidisciplinary collaboration and the management of clinical care pathways for patients with complex healthcare needs.

Flavia Damiano, Surgical Inpatient Unit Nurse Manager, Veneto Institute of Oncology IOV-IRCCS, Padua - Italy

Dr. Flavia Damiano is the Nursing Coordinator for the Oncological Surgery Area at the Veneto Institute of Oncology IOV – IRCCS, Castelfranco Veneto, Italy. In this role, she is involved in the organization and coordination of nursing activities within surgical oncology care settings, contributing to the management of clinical care pathways and supporting multidisciplinary collaboration in an oncology hospital environment.

Cristina Tridello, Surgical Inpatient Unit Nurse Manager, Veneto Institute of Oncology IOV-IRCCS, Padua - Italy

Dr. Flavia Tridello is the Nursing Coordinator for the Inpatient Care Area at the Veneto Institute of Oncology IOV – IRCCS, Castelfranco Veneto, Italy. In this role, she coordinates nursing activities within inpatient wards, contributing to the organization of care processes, the management of clinical pathways and the promotion of multidisciplinary collaboration in an oncology hospital setting.

Marina Coppola, Pharmacy unit, Veneto Institute of Oncology IOV-IRCCS, Padua - Italy

Dr. Marina Coppola is Director of the Hospital Pharmacy Unit and Director of the Department of Translational Oncology and Services at the Veneto Institute of Oncology IOV – IRCCS, Padua, Italy. In these roles, she oversees pharmaceutical services and contributes to the strategic coordination of translational oncology and healthcare service activities within a comprehensive cancer institute.

Most read articles by the same author(s)